CVE-2025-1204
CVE-2025-1204
Título es
CVE-2025-1204
Mar, 25/02/2025 – 17:15
Tipo
CWE-912
Gravedad 2.0 Txt
Pendiente de análisis
Título en
CVE-2025-1204
Descripción en
The "update" binary in the firmware of the affected product sends attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function triggers if the 'C' button is pressed at a specific time during the boot process. If an attacker is able to control or impersonate this IP address, they could upload and overwrite files on the device.
25/02/2025
25/02/2025
Vector CVSS:4.0
CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:P/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X
Gravedad 4.0
7.70
Gravedad 4.0 txt
HIGH
Gravedad 3.1 Txt Gravedad 3.1 (CVSS 3.1 Base Score)
Pendiente de análisis
Referencias
https://claroty.com/team82/research/are-contec-cms8000-patient-monitors-infected-with-a-chinese-backdoor-the-reality-is-more-complicated?ref=vault33.org
https://www.cisa.gov/news-events/ics-medical-advisories/icsma-25-030-01
Enviar en el boletín
Off